ICP 300 IN-CAB PRINTER
9
4 Installation - Digi-Star Indicators
(Printer cable Pt No. S/CB/167-4-007)
Your Digi-Star scale indicator needs to be configured to run with your new
ICP300 printer. Version 8.0 or newer software is required for the printer to
function correctly. A RS232 serial port needs to be present on your scale
indicator, which is normally labeled as “Serial”, “J904”, or “J905” based on
model.
4.1 Verify Software Revision
1. Turn on the scale and wait for normal operation.
2. Press ONto start the self-test sequence (EZ2500V and EZ400/ GT400
models press “NET/GROSS” then “ON/OFF” to start).
3. Watch the self-test sequence on the screen. The words “PRG ID” will be
displayed, followed by the software version code. Press “ONto pause the
test if needed. This software code must be “8.0” or larger to continue;
contact Digi-Star or an authorized service center to upgrade to newer
software if required.
4. Press any other key to cancel the self-test.
4.2 Setup Keypad-based Scales
1. Press 271 on keypad then “Select”.
2. Press “Select” again until “8N1 is displayed.
3. Press “ON” to save.
4. Press 275 on keypad then “Select”.
5. Press “Select” again until “4800” is displayed.
6. Press “ON” to save.
7. Turn off indicator, then turn back on.